Discrepancies between BA, Multifitting and IMD in scattering on roughness
BA gives about 1000 times smaller scattering than IMD and Multifitting for roughness on substrate
Roughness:
- sigma = 0.01 nm
- cor_radius = 100 nm
- hurst = 0.5
Substrate:
- delta = 1e-6, beta = 0
Instrument:
- theta_incident = 0.2 deg
- wavelength = 0.1 nm
- theta_scattered = 0...2 deg
- num_theta_points = 10000
- phi scattered = -1..1 deg
- num_phi_points = 3
- pixel angular size = 4.06e-8 radian
- no footptring
Comparison
- BornAgain: peak scattered intensity (divided by angular pixel size) = 2.45e-6 at theta_scattered = 0.081 deg
- IMD: peak scattered intensity = 3.09e-3 at theta_scattered = 0.081 deg
- Multifitting: peak scattered intensity = 3.09e-3 at theta_scattered = 0.081 deg
This should be investigated further.
IMD:
Multifitting:
BornAgain:
Edited by Mikhail Svechnikov